SP208 is used for sampling of chemical substances in the atmosphere to sample tubes
(a glass tube filled up with packing material) or sampling cartridges.
Features
For accuracy control of flow measurement
SP208 can perform two (2) lines sampling at once by two (2) pumps.
It also can perform integration of suction and its volume.
Flow rate is set to display the value in 20-degree C conversion.
Large flow rate range
1000 type corresponds to the sampling cartridge which suction resistance
can be large
different by formaldehyde and VOC by one set.
100 type is very high accuracy in low flow rate range and suitable for
thermal desorption.
Correspondence tosuction resistance change
The suction capability of pumps is adjusted to be able to perform sampling
at stable flow rate
since it has a function that setting flow rate is maintained even if suction
resistance of trapping
materials or cartridge is changed. Especially when sampling is performed
in a low flow rate range,
the pump's revolution speed is getting slower, and noise or consumption
power can be down.
Multiple functions
A setup of sampling start conditions
Sampling start can be selected from the conditions of Real Time (Year,
Month, Day, hour, and Min.)
after setting time (minute) progress.
A setup of sampling end conditions
Sampling end can be selected from the lapsed time from a sampling start, the total sampling volume
and the conditions of sequencing (it stops manually).
Sampling conditions are saved at file
Checking functions can be performed by a personal computer. Software is an option.
Adjustment of a pulse to be optimal conditions
A pulse style is detected and (High, Good, or Low) is displayed in the 3 color LED to be
the best condition of pump suction capability and suction resistance of a sample tube
or a trapping cartridge.
When pulse (High, Low) is displayed, adjust a pulse valve to be "Good" manually.
Easy to OperateFront View: Gas suction part
・Hose Joint (Standard Accessories)
The Hose Joint can be connected to the soft tubes of 3mm or 5mmID with steps.
・SL Joint
When PTFE tube with SUS piping kit is used, the hose joint will be replaced.

・Filter Joint
The filter can be replaced by customers.
Each joint is connectable.

Back View: Adding a back foot
- The back foot was added to the bottom of the unit.
- SP208II can now be placed on the floor when it is placed while grabbing the handle.
- Detaching the cable is easy since RS232C communication connector is placed in
the front side of the unit.
 
Application
SP208-100Dual II (10-100mL/min.)
- Suitable for sampling (solid phase adsorption / thermal desorption) of VOCs in indoor air measurement.
- Suitable for sampling of building-materials out-gassing with a micro chamber.
- Accuracy guarantee (10-20 mL/min.±10%) in a low flow rate range is offered.
SP208-1000Dual II (100-1000mL/min.)
- Suitable for sampling of the formaldehyde in indoor air measurement, and sampling
(solid phase adsorption / solvent sampling process) of VOCs.
- Suitable also for sampling of SVOC in a habitation residence.
- Suitable for sampling of building-materials out-gassing with a small chamber.
- Suitable for sampling of the ethylene-oxide in the atmosphere and work environment.
Specifications
(Not a CE certified product)
| Model |
SP208-100Dual II |
SP208-1000Dual II |
| Cat.No. |
2702-17576 |
2702-17581 |
| Suction Flow Rate Range |
10 - 100mL/min. |
100 - 1000mL/min. |
| Integrating range |
0.01 - 9999.99L |
| Setting items |
Start Condition : Promptly, After setting time, Real time |
| |
End Condition :After setting time, after reaching to integrating capacity,
Sequence(Manual stop) |
| Storage Files |
5 |
| Data Log Function |
Record up to 30 data. |
| Indicator |
LCD with back light |
| Flow rate fluctuation |
±5% when 24hours operation (under 20℃ stable and within suction flow
rate range) |
| Flow rate Measurement Accuracy |
20-100mL/min. within ±5% |
100mL-1000/min. within ±5% |
| |
10-20mL/min. within ±10% |
100-200mL/min. within ±10% |
| Pulse Display |
High, Good, Lowin 3 colors (LED) |
| Operation Temp. Humidity Range |
DC12V |
| Power supply |
1.2A (DC12v) |
AC100-240V |
| Dimension |
275(W) × 250(D)×100(H)mm |
| Weight |
3.5kg approx. |
Send an inquiry
|