Home > Applications > Automobile Interior VOC Analysis

Automobile Interior VOC Analysis

Products related to Sampling Bag Method

Trapping Set complied with JASO M902:2007; of; Japanese Automobile
Standard

FEATURES

  • Gas Supply unit of 2 lines for N2 gas supply
  • Sampling bag can be cleaned up.
  • Sample can be trapped with 4ea trapping tubes as maximum.
  • 2ea 10L bag or 1ea 20L bag are measurable.
TRAPPING UNIT OVEN UNIT

SAMPLE ANALYSIS EXAMPLES (ACTUAL SAMPLE)

EXAMPLE OF SAMPLING BAG METHOD ANALYSIS
Trapping System :Trapping System
under Sampling Bag Method
Trapping Conditions :65C, 2hr, 1L
Trapping Tube :Tenax TA
TD Conditions :270C, 10min
Split Rate :1/5
COLUMN FLOW SPEED :1mL/min
COLUMN :InertCap 1MS
60m×0.25mm 0.25um
GC CONDITIONS 40C(5min)-280C(21min)
DETECTOR MS SCAN 35~450

EXAMPLE OF STANDARD SAMPLE ANALYSIS W/DNPH CARTRIDGE

1.Formaldehyde 20µg/L
2.Acetoaldehyde 20µg/L
HPLC conditions
Column : Inertsil ODS-SP
(5µm, 150×4.6 mm I.D.)
Eluent A) CH3CN B) H2O
A/B = 55/45, v/v
(gradient mixer)
Flow Volume : 1.0 mL/min
Column Temp. : 40 C
Detector : UV 360nm
Injection Vol. : 20µL

SPECIFICATIONS

Specifications
Oven Program Temperature Program Functions
Temperature Range Room Temp.+10C~210C
Max.Temp.Achi.Speed 50min.(Empty Test)
N2 Gas Supply Flow Vol 0~1000mL/min
Internal Oven Size 450(W)×450(D)×400(H)mm
Dimensions 526(W)×550(D)×795(H)mm
Net Weight 48kg
Power Supply AC100V±10%50/60Hz 13A
Sampling Pump
SP208-1000Dual2
Suction Flow Vol.Range 100~1000mL/min
Dimensions 275(W)×346(D)×100(H)mm
Weight 3.5kg
Power Supply 1.2A(DC12VACAdaptor
Installation Space 900(W)×700(D)mm

>>Next page